Bestman Medical CE ECG/Electrocardiograph ECG-203

Model No.︰

ECG-203

Brand Name︰

BESTMAN

Country of Origin︰

China

Unit Price︰

-

Minimum Order︰

1 pc

Share on:

Product Description

Features:
Digital isolation technology is for minimizing component drift v/s time& temperature to guarantee high-adaptability of the equipment to the environment.

Digital signal processor for effective inhibition of baseline drift, AC interference filter, EMG interference filter and the heart rate, to guarantee the authenticity and dependability.

Auto-regulation of baseline drift can effectively inhibit baseline drift, optimizing the printing position to achieve high-quality ECG.

With a high–resolution thermal printer to print out ECG waveform, describing the trace clear and accurate, annotation as well as related parameters for diagnostic reference.

Roll recording paper for ECG-203 is 63mm in width, simultaneous 12 lead acquisition, and 3 lead live record, high affectivity of ECG examination, good effect and economic utility.

Function of rhythm lead for observing abnormal ECG trace & heart rate.

Supported by AC/Rechargeable battery for continuous examination whenever is necessary. For battery operation, ECG-203 is equipped with a battery charger and a system for battery capacity management and protection.

Safety level for ECG-203 corresponds to Type CF, Class I according to the GB9706.1-1995 criterion.

Specifications:
Lead: Standard 12 leads, changing lead automatically

Input mode: Floating input circuit; Protection against defibrillation effect

leakage current: <10uA

Input Impedance≥50MΩ

Calibration Voltage: 1mV(±3%)

Sampling rate: 1ms

A/D Converter: 12bit

frequency response: 0.05Hz-150Hz(-3dB)

time constant: >3.2s

CMRR: >80dB  >100dB(with filter)

CMRR: >80dB; >100dB(with AC filter in use)

35Hz(-3dB)/25Hz(-3dB)EMG Interference filter: 35Hz(-3dB)/25Hz(-3dB)

Recording Mode: Thermal printer

Paper Speed: 25.50mm/s±3%

Paper Size: 63mmx20m/63mmx30m, roll type: <50nA

input loop current:
DC output:10mm/V,single output port: >100KΩ

DC external input: 10mm/V,single end input: >100KΩ

ECG output:0.5V/mV,single end output: <100Ω

Safety Standard: I-CF, Protector form defibrillation

300mV anti-polarization voltage: <15u Vp-p

Noise Level : <15µVP-P

Power: AC 220V 50Hz/110V 60Hz;  DC 12V

Packaging collocation: 
Single unit Dimension: 41cm2×7.5cm1×7cm

Gross weight: 5kg

Normative collocation: manufacturer's  identify certificate 1 piece, packing list 1 piece, eligibility card 1 piece, warranty card 1 piece,63mm Thermal Recording Paper 1 roll, patient cable 1 piece, Three plugged power cable 1 piece, limb electrode(limb nip) 4 pcs, chest electrode (sucking ball) 6 pcs, AC electric fuse 2 pieces, user manual 1 pcs, Grounding Cable 1 pcs
 